9.1 Introduction
Elements in foods classified according to nutritional essentiality, potential toxic risk, and inclusion in USDA Nutrient Database for Standard Reference
Essential nutrienta |
Toxicity concern |
USDA Nutrient Databaseb |
---|---|---|
Sodium |
Lead |
Calcium |
Potassium |
Mercury |
Iron |
Chloride |
Cadmium |
Magnesium |
Calcium |
Nickel |
Phosphorous |
Chromium |
Arsenic |
Potassium |
Copper |
Thallium |
Sodium |
Fluoride |
Zinc |
|
Iodine |
Copper |
|
Iron |
Manganese |
|
Magnesium |
Selenium |
|
Manganese |
Fluoride |
|
Molybdenum |
||
Phosphorous |
||
Selenium |
||
Zinc |
||
Arsenic |
||
Boron |
||
Nickel |
||
Silicon |
||
Vanadium |
As the names imply, atomic absorption spectroscopy (AAS) quantifies the absorption of electromagnetic radiation by well-separated neutral atoms, while atomic emission spectroscopy (AES) measures emission of radiation from atoms in excited states. AAS and AES allow accurate measurements of mineral elements even in the presence of other components because the atomic absorption and emission spectra are unique for each individual element. The use of inductively coupled plasma (ICP), originally developed in the 1960s [3, 4], as an excitation source for emission spectroscopy has further expanded our ability to rapidly measure multiple elements in a single sample. In theory, virtually all of the elements in the periodic chart can be determined by AAS or AES. In practice, atomic spectroscopy is used primarily to determine mineral elements. Table 8.2 in Chap. 8 shows a comparison of different spectroscopy methods commonly available for food analysis, including AAS and AES.
More recently, ICP has been mated with mass spectrometry (MS) to form ICP-MS instruments that are capable of measuring mineral elements with extremely low detection limits. Moreover, mass spectrometers have the added advantage of being able to separate and quantify multiple isotopes of the same element. Taken together, these instrumental methods have largely replaced traditional wet chemistry methods for food mineral analysis, although traditional methods for calcium, chloride, fluoride, and phosphorus remain in use today (see Chap. 21).
This chapter deals with the basic principles that underlie analytical atomic spectroscopy and provides an overview of the instrumentation available for measuring atomic absorption and emission. A discussion of ICP-MS is also included. Readers interested in a more thorough treatment of the topic are referred to references 5–8.
9.2 General Principles
9.2.1 Energy Transitions in Atoms
9.2.2 Atomization
Methods and temperature ranges for atomization of analytes
Source of energy for atomization |
Approximate atomization temperature range (K) |
Analytical method |
---|---|---|
Flame |
2,000–3,400 |
AAS, AES |
Electrothermal |
1,500–3,300 |
AAS (graphite furnace) |
Inductively coupled argon plasma |
6,000–7,000 |
ICP-OES, ICP-MS |
9.3 Atomic Absorption Spectroscopy
AAS is an analytical method based on the absorption of ultraviolet-visible (UV-Vis) radiation by free atoms in the gaseous state. It is a relatively simple method and was the most widely used form of atomic spectroscopy in food analysis for many years. It has been largely replaced by the more powerful ICP-based spectroscopy. Two types of atomization are commonly used in AAS: flame atomization and electrothermal (graphite furnace) atomization.
9.3.1 Principles of Flame Atomic Absorption Spectroscopy
-
A = absorbance
-
I o = intensity of radiation incident on the flame
-
I = intensity of radiation exiting the flame
-
a = molar absorptivity
-
b = path length through the flame
-
c = concentration of atoms in the flame
Clearly, absorbance is directly related to the concentration of atoms in the flame.
9.3.2 Principles of Electrothermal (Graphite Furnace) Atomic Absorption Spectroscopy
Electrothermal AAS is identical to flame AAS except for the atomization process. In electrothermal AAS, the sample is heated electrically in stages inside a graphite tube, commonly known as graphite furnace, to achieve atomization. The tube is aligned to the path of the radiation beam to be absorbed by the atomized sample and absorbance is determined. Electrothermal AAS requires smaller sample sizes and offers lower detection limits. Disadvantages are the added expense of the graphite furnace, lower sample throughput, higher matrix interferences, and lower precision.
9.3.3 Instrumentation for Atomic Absorption Spectroscopy
- 1.
Radiation source, a hollow cathode lamp (HCL) or an electrode-less discharge lamp (EDL)
- 2.
Atomizer, usually a nebulizer-burner system or a graphite furnace
- 3.
Monochromator, usually an UV-Vis grating monochromator
- 4.
Detector, a photomultiplier tube (PMT) or a solid-state detector (SSD)
- 5.
Readout device, an analog or a digital readout
(Radiation source and atomizers will be further discussed in the following paragraphs. See Chap. 7, Sects. 7.2.6.2, 7.2.6.3, and 7.2.6.4 for a more detailed discussion of monochromators, detectors, and readout devices, respectively.)
9.3.3.1 Radiation Source
An electrode-less discharge lamp (EDL) contains no electrodes but a hollow glass vessel filled with an inert gas plus the element of interest. The discharge is produced by a high-frequency generator coil rather than an electric current [9]. EDLs are suitable for volatile elements such as arsenic, mercury, and cadmium.
Radiation reaching the monochromator comes from three sources: (1) attenuated beam from the HCL (specific emission), (2) emission from sample atoms (including both analyte and non-analyte atoms) that were excited by the flame (nonspecific emission), and (3) radiation resulting from the combustion of the fuel to create the flame (nonspecific emission). Instruments are designed to eliminate nonspecific emissions from reaching the detector. This is accomplished by positioning a monochromator between the flame and the detector. The monochromator disperses wavelengths of light that are not specific to the analyte element and isolates a line that is specific. Thus, radiation reaching the detector is the sum of radiation from the attenuated HCL beam and radiation emitted by excited analyte atoms in the flame. Since we are interested only in the amount of HCL radiation absorbed by analyte atoms in the flame, it is necessary to correct for emission from excited analyte atoms in the flame. This is accomplished by positioning a beam chopper perpendicular to the light path between the HCL and the flame (Fig. 9.3). A beam chopper is a disk with segments removed. The disk rotates at a constant speed so that the light emitted from the HCL reaching the detector is either unimpeded or blocked at regular intervals, i.e., it is alternating. In contrast, emission from excited analyte atoms in the flame reaching the detector is continuous. The instrument electronics subtracts the continuous emission signal from the alternating emission signal so only the signal from the attenuated HCL beam is recorded in the readout.
9.3.3.2 Atomizers
Flame and graphite furnace atomizers are the two common types of atomizers used in AAS. When applicable, a cold vapor technique for mercury and a hydride generation technique for a few elements are used to enhance sensitivity.
The flame atomizer consists of a nebulizer and a burner. The nebulizer is designed to convert the sample solution into a fine mist or aerosol. This is accomplished by aspirating the sample through a capillary into a chamber through which oxidant and fuel are flowing. The chamber contains baffles that remove larger droplets, leaving a very fine mist. Only about 1 % of the total sample is carried into the flame by the oxidant-fuel mixture. The larger droplets fall to the bottom of the mixing chamber and are collected as waste. The burner head contains a long, narrow slot that produces a flame that may be 5–10 cm in length. This gives a long path length that increases the sensitivity of the measurement.
- 1.
Stoichiometric. This flame is produced from stoichiometric amounts (exact reaction ratios) of oxidant and fuel, so the fuel is completely burned and the oxidant is completely consumed. It is characterized by yellow fringes.
- 2.
Oxidizing. This flame is produced from a fuel-lean (excess of oxygen) mixture. It is the hottest flame and has a clear blue appearance.
- 3.
Reducing. This flame is produced from a fuel-rich mixture (excess of fuel compared to oxygen). It is a relatively cool flame and has a yellow color.
Analysts should follow manufacturer’s guidelines or consult the literature for the proper type of flame for each element.
Flame atomizers have the advantage of being stable and easy to use. However, sensitivity is relatively low because much of the sample never reaches the flame and the residence time of the sample in the flame is short.
The graphite furnace is typically a cylindrical graphite tube connected to an electrical power supply. The sample is injected into the tube through an inlet using a microliter syringe with sample volumes ranging from 0.5 to 100 μL. During operation, the system is flushed with an inert gas to prevent the tube from burning and to exclude air from the sample compartment. The tube is heated electrically in stages: first the sample solvent is evaporated, then the sample is ashed, and finally the temperature is rapidly increased to ~2000–3000 K to quickly vaporize and atomize the sample.
The cold vapor technique works only for mercury, because mercury is the only mineral element that can exist as free atoms in the gaseous state at room temperature. Mercury compounds in a sample are first reduced to elemental mercury by the action of stannous chloride, a strong reducing agent. The elemental mercury is then carried in a stream of inert gas into an absorption cell without the need for further atomization. The hydride generation technique is limited to elements capable of forming volatile hydrides that include arsenic, lead, tin, bismuth, antimony, tellurium, germanium, and selenium. Samples containing these elements are reacted with sodium borohydride to generate volatile hydrides, which are carried into an absorption cell and decomposed by heat. Absorbance measurements with these two techniques are conducted in the same manner as with flame atomization, but sensitivity is greatly enhanced because there is very little sample loss [5].
9.3.4 General Procedure for Atomic Absorption Analysis
While the basic design of all atomic absorption spectrometers is similar, operation procedures do vary from one instrument to another. For any given method, it is always a good practice to carefully review standard operating procedures provided by the manufacturer before using the instrument. Most instruction manuals also provide pertinent information for the analysis of each particular element (wavelength and slit width requirements, interferences and corrections, flame characteristics, linear ranges, etc.).
9.3.4.1 Safety Precautions
General laboratory safety protocols and procedures as well as safety precautions recommended by the instrument manufacturers must be followed carefully to avoid personal injuries or costly accidents. The most commonly used fuel sources in flame AAS are mixtures of air-acetylene and nitrous oxide-acetylene. ACETYLENE IS AN EXPLOSIVE GAS. Proper ventilation must be in place before operation. The exhaust vent should be positioned directly above the burner to avoid the buildup of unburned fuel or any potentially hazardous toxic fumes. Flame atomic absorption spectrometers should never be left unattended while in operation.
9.3.4.2 Calibration
Instruction manuals from the manufacturers may provide values for characteristic concentrations for each element. For example, manuals for Perkin-Elmer atomic absorption spectrophotometers state that a 5.0 mg/L aqueous solution of iron “will give a reading of approximately 0.2 absorbance units.” If the measured absorbance reading deviates significantly from this value, appropriate adjustments (e.g., flame characteristics, lamp alignment, etc.) should be made.
9.3.5 Interferences in Atomic Absorption Spectroscopy
Two types of interferences are encountered in AAS: spectral and nonspectral. Spectral interferences are caused by the absorption of radiation by other elemental or molecular species at wavelengths that overlap with the spectral regions of the analyte present in the sample. Nonspectral interferences are caused by sample matrices and conditions that affect the atomization efficiency and/or the ionization of neutral atoms in the atomizer.
9.3.5.1 Spectral Interferences
An element in the sample other than the element of interest may absorb at the wavelength of the spectral band being used. Such interference is rare because emission lines from the HCL are so narrow that only the element of interest is capable of absorbing the radiation in most cases. One example of when this problem does occur is with the interference of iron in zinc determinations. Zinc has a spectral line at 213.856 nm, which overlaps the iron line at 213.859 nm. The problem may be solved by choosing an alternative spectral line for measuring zinc or by narrowing the monochromator slit width.
The presence of alkaline earth oxide and hydroxide molecules may also lead to several specific spectral interferences. Spectra of calcium oxide and magnesium hydroxide will appear as background absorption for atomic absorption measurements of sodium and chromium, respectively [10]. These interferences are weak but must be taken into account when working with a complex sample matrix.
9.3.5.2 Nonspectral Interferences
As mentioned above, quantitative results for an unknown sample are possible only through comparison with a series of standards of known concentrations. Transport interferences may occur when other components present in the sample matrix influence physical properties such as viscosity, surface tension, and vapor pressure of the sample solution, leading to differences in the rate of aspiration, nebulization, or transport between the sample solution and the standards during flame atomization. Such interferences often can be overcome by using the same solvent and by matching as closely as possible the physical properties of the sample solution and the standards. The standard addition protocol (see Chap. 7, Sect. 7.2.4) may also be used. Transport interferences are rarely a problem with graphite furnace instruments but matrix interferences are a common and serious problem.
Matrix composition of the sample solution also may affect the lateral migration of an analyte, resulting in solute vaporization interferences. For example, it has been observed in flame absorption and emission that alkaline earth metals are depressed by elevated levels of aluminum and phosphorus [11], and aluminum also suppresses the recovery of calcium [12]. Chemical interferences occur when an element forms thermally stable compounds that do not decompose during atomization. Refractory metals such as titanium and molybdenum may combine with oxygen to form stable oxides; higher-temperature flames are usually required for their dissociation. Also, phosphate in a sample matrix reacts with calcium to form calcium pyrophosphate which is not decomposed in the flame. A releasing agent such as lanthanum, which binds phosphate more strongly than calcium, may be added to the sample solution and the standards to free up calcium for atomization [12].
9.4 Atomic Emission Spectroscopy
Energy for excitation may be produced by heat (usually from a flame), light (from a laser), electricity (arcs or sparks), or radio waves (ICP). (Note: AES is also commonly called optical emission spectroscopy (OES), especially when combined with ICP. In this chapter, we will use ICP-OES rather than ICP-AES for our discussion, but the two terms are virtually interchangeable.)
The two most common forms of AES used in food analysis are flame emission spectroscopy and inductively coupled plasma-optical emission spectroscopy (ICP-OES).
9.4.1 Principles of Flame Emission Spectroscopy
Flame emission spectrometers employ a nebulizer-burner system to atomize and excite the atoms of the elements being measured. The flame with the excited atoms serves as the radiation source, so an external source (the HCL with the beam chopper) is not required. Otherwise instrumentation for flame emission spectroscopy is essentially identical to that for AAS. Many modern atomic absorption spectrometers can also be operated as flame emission spectrometers.
In some instruments, interference filters (instead of the more versatile monochromators typically found in absorption/emission spectrometers) are employed to isolate the spectral region of interest. Flame photometers are economical emission spectrometers equipped with interference filters and are specifically designed for the analysis of the alkali and alkaline earth metals in biological samples. Low flame temperatures are used so that only easily excited elements such as sodium, potassium, and calcium produce emissions. This results in a simpler spectrum and reduces interference from other elements present in the sample matrix.
9.4.2 Principles of Inductively Coupled Plasma-Optical Emission Spectroscopy
ICP-OES differs from flame emission spectroscopy in that it uses an argon plasma as the excitation source. A plasma is defined as a gaseous mixture containing significant concentrations of cations and electrons. Temperatures in argon plasmas could be as high as 10,000 K, with analyte excitation temperature typically ranging from 6,000 to 7,000 K.
The extremely high temperatures and the inert atmosphere of argon plasmas are ideal for the atomization, ionization, and excitation of the analyte atoms in the sample. The low oxygen content reduces the formation of oxides, which is sometimes a problem with flame methods. The nearly complete atomization of the sample minimizes chemical interferences. The relatively uniform temperatures in plasmas (compared to nonuniform temperatures in flames) and the relatively long residence time give good linear responses over a wide concentration range (up to 6 orders of magnitude).
9.4.3 Instrumentation for Inductively Coupled Plasma-Optical Emission Spectroscopy
- 1.
Argon plasma torch
- 2.
Monochromator, polychromator, or echelle optical system
- 3.
Detector(s), a single or multiple PMT(s) or solid-state array detector(s)
- 4.
Computer for data collection and treatment
9.4.3.1 Argon Plasma Torch
9.4.3.1.1 Characteristics of an Argon Plasma Torch
9.4.3.1.2 Sample Introduction and Analyte Excitation
Samples are nebulized and introduced as aerosols carried by another stream of argon gas in the inner tube inside the annulus of the plasma at the base of the RF load coil. The sample goes through the process of desolvation, vaporization, atomization, ionization, and excitation as shown in Fig. 9.2. The exact mechanisms by which the analyte atoms and ions are excited in a plasma are not understood fully. Nevertheless, there is general agreement that excitation is dependent primarily on the number and temperature of the electrons in the plasma [14]. Presumably, when electrons are accelerated in a magnetic field, they acquire enough kinetic energy to excite analyte atoms and ions upon collision [14]. Exceptions to this mechanism include the excitation of neutral atoms of magnesium, copper, and a few other elements that are believed to be excited when an argon ion (Ar+) extracts an electron from the analyte atom (M) to yield M+* and Ar0, where M is a generic abbreviation for a ground-state metal atom and M+* is an excited metal ion. This mechanism is termed “charge transfer” [15, 16].
9.4.3.1.3 Radial and Axial Viewing
9.4.3.2 Detectors and Optical Systems
The echelle optical system employs two dispersing components in series, a prism and a diffraction grating. The prism first disperses the radiation from the plasma torch without any wavelength overlap (in the x-direction). The radiation then strikes a low-density ruled grating (about 53 groves per mm). This further separates the radiation in a direction perpendicular to the direction of radiation dispersed by the prism (in the y-direction), producing a two-dimensional spectrum with a wavelength range of 166–840 nm. When the radiation passing through the echelle optical system is focused onto the solid-state array detector, electrons are liberated proportionally to the intensity of the incident radiation and trapped in the silicon-based, light-sensitive elements called pixels for signal processing. ICP-OES instruments typically use one of three types of solid-state array detectors: charge coupled device (CCD), complementary metal oxide semiconductor (CMOS), or charge injection device (CID). A description of these detectors is beyond the scope of this chapter. Interested readers are referred to the comparisons between different types of solid-state array detectors provided in references [17, 18].
9.4.4 General Procedure for Inductively Coupled Plasma-Optical Emission Spectroscopy Analysis
As is the case with atomic absorption spectrometers, operating procedures for atomic emission spectrometers vary somewhat from instrument to instrument. ICP-OES instruments are almost always interfaced with computers. The software contains methods that specify instrument operating conditions. The computer may be programmed by the operator, or in some cases, default conditions may be used. Once the method is established, operation is highly automated.
9.4.5 Interferences in Inductively Coupled Plasma-Optical Emission Spectroscopy
9.5 Applications of Atomic Absorption and Emission Spectroscopy
9.5.1 Uses
Atomic absorption and emission spectroscopy are widely used for the quantitative measurement of mineral elements in foods. In principle, any food may be analyzed with any of the atomic spectroscopy methods discussed. In most cases, it is necessary to ash the food to destroy organic matter and to dissolve the ash in a suitable solvent (usually water or dilute acid) prior to analysis (see Chap. 16 for details on ashing methodology). Proper ashing is critical to accuracy. Some elements may be volatile at temperatures used in dry ashing procedures. Volatilization is less of a problem in wet ashing, except for the determination of boron, which is recovered better using a dry ashing method. However, ashing reagents may be contaminated with the analyte. It is therefore wise to carry blanks through the ashing procedure.
Some liquid products may be analyzed without ashing, provided appropriate precautions are taken to avoid interferences. For example, vegetable oils may be analyzed by dissolving the oil in an organic solvent such as acetone or ethanol and aspirating the solution directly into a flame atomic absorption spectrometer. Milk samples may be treated with trichloroacetic acid to precipitate the protein; the resulting supernatant is analyzed directly. A disadvantage of this approach is that the sample is diluted in the process and the analyte can become entrapped or complexed to the precipitated proteins. This may be a problem when analytes are present in low concentrations. An alternative approach is to use a graphite furnace for atomization. For example, an aliquot of an oil may be introduced directly into a graphite furnace for atomization. The choice of method will depend on several factors, including instrument availability, cost, precision/sensitivity, and operator skill.
9.5.2 Practical Considerations
9.5.2.1 Reagents
Since concentrations of many mineral elements in foods are at the trace level, it is essential to use highly pure chemical reagents and water for preparation of samples and standard solutions. Only reagent grade chemicals should be used. Water may be purified by distillation, deionization, or a combination of the two. Reagent blanks should always be carried through the analysis.
9.5.2.2 Standards
Quantitative atomic spectroscopy depends on comparison of the sample measurement with appropriate standards. Ideally, standards should contain the analyte metal in known concentrations in a solution that closely approximates the sample solution in composition and physical properties. Because many factors can affect the measurement, such as flame temperature, aspiration rate, and the like, it is essential to run standards frequently, preferably right before and/or right after running the sample. Standard solutions may be purchased from commercial sources, or they may be prepared by the analyst. Obviously, standards must be prepared with extreme care since the accuracy of the analyte determination depends on the accuracy of the standard. Perhaps the best way to check the accuracy of a given assay procedure is to analyze a reference material of known composition and similar matrix. Standard reference materials may be purchased from the United States National Institute of Standards and Technology (NIST) [19].
9.5.2.3 Labware
Vessels used for sample preparation and storage must be clean and free of the elements of interest. Plastic containers are preferable because glass has a greater tendency to adsorb and later leach metal ions. All labware should be thoroughly washed with a detergent, carefully rinsed with distilled or deionized water, soaked in an acid solution (1 N HCl is sufficient for most applications), and rinsed again with distilled or deionized water.
9.6 Inductively Coupled Plasma-Mass Spectrometry
9.6.1 Principles of Inductively Coupled Plasma-Mass Spectrometry
The principles of mass spectrometry and descriptions of instrumentation for different types of mass spectrometers are given in detail in Chap. 11. In ICP-MS mineral analyses, samples are prepared and aspirated into the ICP torch in the same manner as in ICP-OES, but instead of having an optical system and a device for separating and detecting radiation of specific wavelengths, the ICP-MS uses a mass spectrometer to separate and detect ions of the elements directly based on their unique mass-to-charge (m/z) ratios. As shown in Fig. 9.15, the interface between the high-temperature ICP operating at atmospheric pressure and the high vacuum mass spectrometer consists of: (1) two funnel-shaped water-cooled cones (the sampling cone and the skimmer cone) and (2) ion lenses (extraction lenses and omega lenses) to guide the analyte ions into the quadrupole while removing electrons, photons, and other neutral species from the ICP discharge.
9.6.2 Interferences in Inductively Coupled Plasma-Mass Spectrometry
Interferences in ICP-MS arise when different ionic species from the sample have the same m/z ratio, leading to overlapping of signals. For example, iron has four naturally occurring stable isotopes: 54Fe, 56Fe, 57Fe, and 58Fe, with natural abundances of 5.8 %, 91.75 %, 2.1 %, and 0.28 %, respectively. Nickel has five stable isotopes: 58Ni, 60Ni, 61Ni, 62Ni, and 64Ni, with natural abundances of 68.04 %, 26.22 %, 1.14 %, 3.63 %, and 0.93 %, respectively. The signals for 58Fe and 58Ni will overlap, resulting in an isobaric interference because m/z = 58 for both isotopes. In this case, the analyst would select 56Fe for the determination of iron and 60Ni for the determination of nickel in the sample. (It is best to select the isotope with the highest natural abundance because the measurement precision is higher for more abundant isotopes. The concentration of the isotope in the sample is equal to the concentration of the element multiplied by the % natural abundance.) Most elements have at least one isotope with a unique mass number, thus allowing identification and quantification of elements.
Examples of polyatomic interference in ICP-MS
Polyatomic species |
m/z |
Interfered element/isotope |
---|---|---|
38Ar1H+ |
39 |
39K+ |
35Cl16O+ |
51 |
51V+ |
40Ar12C+ |
52 |
52Cr+ |
38Ar16O1H+ |
55 |
55Mn+ |
40Ar16O+ |
56 |
56Fe+ |
40Ar35Cl+ |
75 |
75As+ |
40Ar40Ar+ |
80 |
80Se+ |
Another approach to reduce interferences from polyatomic species is to use a high-resolution ICP-MS, which utilizes a double focusing sector-field mass spectrometer to separate ions generated by the plasma [20]. For example, high-resolution ICP-MS has been shown to resolve the Fe peak at mass 55.935 from ArO at mass 55.957 [21, 22]. In this case the ions are measured sequentially, which is the case for most of the ICP-MS systems. However there is one ICP-MS system available that is capable of measuring all the elements from lithium to uranium simultaneously [23]. This system uses a double focusing sector-field mass spectrometer in a Mattauch-Herzog geometry, in which the ion beam energy band width is reduced using an electrostatic energy analyzer to achieve high resolution. The ion beam then passes through a magnetic field for mass separation and is focused on to one focal plane, enabling the entire spectrum to be measured with a flat surface detector simultaneously. This would be an ideal instrument for measuring isotope ratios from transient signals.
9.7 Comparison of AAS, ICP-OES, and ICP-MS
Advantages and disadvantages of AAS, ICP-OES and ICP-MS
Flame AAS |
Graphite furnace AAS |
ICP-OES |
ICP-MS |
|
---|---|---|---|---|
Detection limita |
Good detection limits with many elements at the part per billion (ppb) level |
Better than flame AAS and better than ICP-OES for some elements |
Better than flame AAS |
Overall best detection limits compared to other techniques |
Elemental analytical capability |
Single |
Single |
Multiple |
Multiple |
Approximate analytical working range |
3 orders of magnitude |
2 orders of magnitude |
6 orders of magnitude (could be higher with dual-view models) |
9 orders of magnitude |
Cost |
Low |
Low to medium |
Medium |
High |
Use of explosive fuel gas |
Yes (Flame AAS instruments should not be unattended while in operation.) |
No |
No |
No |
User-friendliness |
Some skills required but relatively easy to use |
Some skills required |
Easy to use once the computer interface is set up and operation is automated |
Method development requires more expertise compared to other techniques |
Ideal application |
Analyzing a limited number of elements in a given sample |
Analyzing a limited number of elements, and requiring better detection limits than Flame AAS |
Multiple elements in a large number of samples |
Multiple elements at ultra-trace concentrations in a large number of samples |
Isotopic analysis |
N/A |
N/A |
N/A |
Isotopic analysis possible because isotopes of the same element have different mass-to-charge ratios |
ICP-OES instruments are capable of determining concentrations of multiple elements in a single sample with a single aspiration. This offers a significant speed advantage and higher throughput over AAS when the objective is to quantify multiple elements (up to 70) in a given sample. The high temperature of the ICP torch also eliminates many nonspectral interferences (e.g., chemical interferences) encountered in AAS. Another advantage of ICP-OES over AAS is a much wider analytical working range. The analytical working range for ICP-OES is 4–6 orders of magnitude (i.e., 1 μg/L to 1 g/L without having to recalibrate the instrument), compared to about 3 orders of magnitude for AAS (i.e., 1 μg/L to 1 mg/L). All these advantages help explain the popularity of ICP-OES in commercial laboratories analyzing multiple elements in a large number of samples. ICP-OES is commonly used to obtain information for standard nutrition labeling.
ICP-MS retains the advantages offered by ICP-OES but, in conjunction with mass spectrometry, offers the lowest detection limits, enhanced multielement capabilities, a wider analytical working range (9 orders of magnitude, i.e., 1 ng/L to 1 g/L), and isotopic information potentially for tracking the geographical origins of food products [25]. Laboratories analyzing trace or ultra-trace concentrations of toxic heavy metals such as cadmium would be best served with an ICP-MS system. The major disadvantage for ICP-MS is perhaps the cost, which is about two to four times higher than its ICP-OES counterparts.
Approximate detection limits (μg/L) for selected elements analyzed with various instruments
Element |
Flame AAS |
Graphite furnace AAS |
ICP-OES |
ICP-MS |
---|---|---|---|---|
Al |
45 |
0.1 |
1 |
0.0004 |
As |
150 |
0.05 |
1 |
0.0004 |
Ca |
1.5 |
0.01 |
0.05 |
0.0003 |
Cd |
0.8 |
0.002 |
0.1 |
0.00007 |
Cu |
1.5 |
0.014 |
0.4 |
0.0002 |
Fe |
5 |
0.06 |
0.1 |
0.0005 |
K |
3 |
0.005 |
1 |
0.001 |
Hg |
300 |
0.6 |
1 |
0.001 |
Mg |
0.15 |
0.004 |
0.04 |
0.0001 |
Mn |
1.5 |
0.005 |
0.1 |
0.0001 |
Na |
0.3 |
0.005 |
0.5 |
0.0003 |
Ni |
6 |
0.07 |
0.5 |
0.0002 |
P |
75,000 |
130 |
4 |
0.04 |
Pb |
15 |
0.05 |
1 |
0.00004 |
Se |
100 |
0.05 |
2 |
0.0003 |
Tl |
15 |
0.1 |
2 |
0.00001 |
Zn |
1.5 |
0.02 |
0.2 |
0.0007 |
9.8 Summary
In comparison with traditional wet chemistry methods, AAS, AES, and ICP-MS methods are capable of measuring trace concentrations of elements in complex matrices rapidly and with excellent precision. For most applications, sample preparation involves the destruction of organic matter by dry or wet ashing, followed by dissolution of the ash in an aqueous solvent, usually a dilute acid. The sample solution is introduced as a fine mist into a flame atomizer or an ICP torch (or by injection into a graphite furnace) where it encounters very high temperatures (~2000–3000 K for flame or graphite furnace, and ~6000–7000 K for plasma). The sample goes through the process of desolvation, vaporization, atomization, and ionization. Analyte atoms, now in the gaseous state, are well separated and remain mostly neutral in a flame, but a significant fraction of them lose an electron and become charged in a plasma. The final step is to measure quantitatively the concentrations of the elements either by atomic spectroscopy or mass spectrometry.
Atomic spectroscopy depends on the absorption or emission of electromagnetic radiation (light) by the atoms in the gaseous state. Atoms absorb or emit radiation of discrete wavelengths because the allowed energy levels of electrons in atoms are fixed and distinct. In other words, each element has a unique set of allowed electronic transitions and therefore a unique spectrum. In AAS, light of a discrete wavelength from the element-specific hollow cathode lamp will only be absorbed by the atoms of that element in the sample. Furthermore, the amount of light absorbed is directly related to the concentration of the atoms in the sample. By measuring the absorbance of light of a particular wavelength by an atomized sample, analysts can determine the concentration of an element even in the presence of other elements. In emission spectroscopy, the optical approach involves exciting the electrons in an element to a higher-energy state by a flame or plasma, and then measuring the intensity of the light emitted when the electrons fall back to the ground state or a lower-energy state. Emission spectroscopy instruments are designed to separate the light emitted from excited atoms and to quantitatively measure the intensity of the emitted light.
In contrast to atomic spectroscopy, ICP-MS instruments are designed to measure ions of the element directly. This necessitates the ionization of atoms in the plasma. The ions of the element are then guided into a mass spectrometer which separates and detects ions according to their unique mass-to-charge (m/z) ratio. Quantification of elements with high sensitivity and specificity can be achieved because most elements have at least one isotope with a unique mass number.
Atomic spectroscopy is a powerful tool for the quantitative measurement of elements in foods. The development of these technologies over the past six decades has had a major impact on food analysis. Today, accurate and precise measurements of a large number of mineral nutrients and non-nutrients in foods can be made rapidly using commercially available instrumentation. Analysts contemplating what instruments to acquire could make a decision based on an assessment of the required cost, user-friendliness, analytical working range, detection limit, multielement capability, and the need for isotopic data.
9.9 Study Questions
- 1.
AAS and AES instruments rely on energy transitions in atoms of elements being measured. What is an “energy transition” in this context and why can it be used to detect and quantify a given element in a sample containing multiple elements? What is the source of energy that produces this energy transition in an AAS instrument? In an AES instrument?
- 2.
Describe the process of “atomization” as it pertains to AAS and AES analyses.
- 3.
Your boss wants to purchase an AAS instrument for your analytical laboratory because it is cheaper but you want an ICP-OES instrument because it is more versatile and will greatly increase your sample throughput. To convince your boss to go with the ICP-OES, you need to educate him on the capabilities and operating principles of the two instruments. Keep in mind that your boss is a food scientist who has not worked in a lab in 20 years.
- (a)
Explain the underlying principles of operation for an ICP-OES instrument in language your boss can understand. Describe the instrument you want to purchase (a simple drawing of the instrument might be helpful here).
- (b)
Explain how AAS differs in instrumentation and principle of operation from what you described previously for ICP-OES.
- (c)
Can you make the case that costs for an ICP-OES would be lower over the long term?
- (d)
For most types of food samples other than clear liquids, what type of sample preparation and treatment is generally required before using ICP-OES or AAS for analysis?
- (a)
- 4.
You are training a new technician in your analytical laboratory on mineral analysis by AAS and ICP-OES. Briefly describe the purpose of each of the following items
- (a)
HCL in AAS
- (b)
Plasma in ICP-OES
- (c)
Echelle optical system in ICP-OES
- (d)
Nebulizer in AAS and ICP-OES
- (a)
- 5.
In the quantitation of Na by atomic absorption, KCl or LiCl was not added to the sample. Would you likely over- or underestimate the true Na content? Explain why either KCl or LiCl is necessary to obtain accurate results.
- 6.
Give five potential sources of error in sample preparation prior to atomic absorption analysis.
- 7.
You are performing iron analysis on a milk sample using AAS. Your results for the blank are high. What could be causing this problem and what is a possible remedy?
- 8.
The detection limit for calcium is lower for ICP-OES than it is for flame AAS. How is the detection limit determined, and what does it mean?
- 9.
When analyzing a sample for mineral elements using ICP-MS, the instrument is programmed to count the number of ions with a specific m/z ratio striking the detector. You decide to determine the concentrations of potassium and calcium in a sample of wheat flour. What m/z ratio would you use for potassium? For calcium? Why? (Hint, study the masses of all the naturally occurring and stable isotopes for the two elements and for argon (see table) and select isotopes with no interferences.) Why is it important to know the masses of argon isotopes as well as potassium and calcium?
Isotope |
Natural abundance (%) |
---|---|
36Ar |
0.34 |
38Ar |
0.063 |
40Ar |
99.6 |
39K |
93.2 |
40K |
0.012 |
41K |
6.73 |
40Ca |
96.95 |
42Ca |
0.65 |
43Ca |
0.14 |
44Ca |
2.086 |
46Ca |
0.004 |
9.10 Practice Problems
- 1.
Your company manufactures and markets an enriched all-purpose flour product. You purchase a premix containing elemental iron powder, riboflavin, niacin, thiamin, and folate which you mix with your flour during milling. To comply with the standard of identity (see Chap. 2) for enriched flour, you specified to the supplier that the premix be formulated so that when added to flour at a specified rate, the concentration of added iron is 20 mg/lb flour. However, you have reason to believe that the iron concentration in the premix is too low so you decide to analyze your enriched flour using your new atomic absorption spectrometer. You follow the following protocol to determine the iron concentration.
- (a)
Weigh out 10.00 g of flour, in triplicate (each replicate should be analyzed separately).
- (b)
Transfer the flour to an 800-mL Kjeldahl flask.
- (c)
Add 20 mL of deionized water, 5 mL of concentrated H2SO4, and 25 mL of concentrated HNO3.
- (d)
Heat on a Kjeldahl burner in a hood until white SO3 fumes form.
- (e)
Cool, add 25 mL of deionized water, and filter quantitatively into a 100-mL volumetric flask. Dilute to volume.
- (f)
Prepare iron standards with concentrations of 2, 4, 6, 8, and 10 mg/L.
- (g)
Install an iron hollow cathode lamp in your atomic absorption spectrometer and turn on the instrument and adjust it according to instructions in the operating manual.
- (h)
Run your standards and each of your ashed samples and record the absorbances.
Calculate the iron concentration in each of your replicates, express as mg Fe/lb flour.Absorbance data for iron standards and flour samples
Sample
Fe Conc. (mg/L)
Absorbance
Corrected absorbance
Reagent blank
–
0.01
–
Standard 1
2.0
0.22
0.21
Standard 2
4.0
0.40
0.39
Standard 3
6.0
0.63
0.62
Standard 4
8.0
0.79
0.78
Standard 5
10.0
1.03
1.02
Flour sample 1
?
0.28
0.27
Flour sample 2
?
0.29
0.28
Flour sample 3
?
0.26
0.25
- (a)
- 2.
Describe a procedure for determining calcium, potassium, and sodium in infant formula using ICP-OES. Note: Concentrations of Ca, K, and Na in infant formula are around 700 mg/L, 730 mg/L, and 300 mg/L, respectively.
Answers
- 1.
The following steps may be used to determine the iron concentration in the flour samples.
- (a)
Enter the data for the standards into Excel. Using the scatter plot function, plot the standard curve and generate a trend line using linear regression. Include the equation for the line and the R2 value. Your results should look like the standard curve shown.
- (b)
Using the equation, calculate the iron concentration in the solution in the volumetric flask for each of your samples. Your answers should be 2.68 mg/L, 2.79 mg/L, and 2.48 mg/L for samples 1, 2, and 3, respectively. The mean is 2.65 mg/L; the standard deviation is 0.16.
- (c)
Now determine the iron concentration in the flour. Recall that you transferred the solution from the Kjeldahl flask quantitatively into the 100-mL volumetric flask. Therefore all of the iron in the flour sample should be in the volumetric flask. The mean concentration is 2.65 mg/L. The volume is 0.1 L. Therefore, the amount of iron in the 10 g of flour is 0.265 mg. To convert this to mg/lb, multiply by 454/10: 0.265 mg/10 g × 454 g/lb = 12 mg Fe/lb flour
- (d)
Your suspicions are confirmed; your supplier shorted you on iron in the premix. You need to correct this as soon as possible because your flour does not conform to the FDA’s standard of identity for enriched flour and you may be subject to legal action by the FDA.
- (a)
- 2.
Consult AOAC Method 984.27 (see Chap. 1 for a description of AOAC International), and the following approach may be used:
- (a)
Shake can vigorously.
- (b)
Transfer 15.0 mL of formula to a 100-mL Kjeldahl flask. (Carry two reagent blanks through with sample.)
- (c)
Add 30 mL of HNO3-HClO4 (2:1).
- (d)
Leave samples overnight.
- (e)
Heat until ashing is complete (follow AOAC procedure carefully – mixture is potentially explosive.)
- (f)
Transfer quantitatively to a 50-mL vol flask. Dilute to volume.
- (g)
Calibrate instrument. Choose wavelengths of 317.9 nm, 766.5 nm, and 589.0 nm for Ca, K, and Na, respectively. Prepare calibration standards containing 200 μg/mL, 200 μg/mL, and 100 μg/mL for Ca, K, and Na, respectively.
- (h)
The ICP-OES computer will calculate concentrations in the samples as analyzed. To convert to concentrations in the formula, use the following equation:
- (a)
Open Access This chapter is licensed under the terms of the Creative Commons Attribution-NonCommercial 2.5 International License (http://creativecommons.org/licenses/by-nc/2.5/), which permits any noncommercial use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license and indicate if changes were made.
The images or other third party material in this chapter are included in the chapter's Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the chapter's Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.