7.1 Introduction
Spectroscopy in the ultraviolet-visible (UV-Vis) range is one of the most commonly encountered laboratory techniques in food analysis. Diverse examples, such as the quantification of macrocomponents (total carbohydrate by the phenol-sulfuric acid method), quantification of microcomponents (thiamine by the thiochrome fluorometric procedure), estimates of rancidity (lipid oxidation status by the thiobarbituric acid test), and surveillance testing (enzyme-linked immunoassays), are presented in this text. In each of these cases, the analytical signal for which the assay is based is either the emission or absorption of radiation in the UV-Vis range. This signal may be inherent in the analyte, such as the absorbance of radiation in the visible range by pigments, or a result of a chemical reaction involving the analyte, such as the colorimetric copper-based Lowry method for the analysis of soluble protein.
Spectrum of visible radiation
Wavelength (nm) |
Color |
Complementary huea |
---|---|---|
<380 |
Ultraviolet |
|
380–420 |
Violet |
Yellow-green |
420–440 |
Violet-blue |
Yellow |
440–470 |
Blue |
Orange |
470–500 |
Blue-green |
Red |
500–520 |
Green |
Purple |
520–550 |
Yellow-green |
Violet |
550–580 |
Yellow |
Violet-blue |
580–620 |
Orange |
Blue |
620–680 |
Red |
Blue-green |
680–780 |
Purple |
Green |
>780 |
Near infrared |
7.2 Ultraviolet and Visible Absorption Spectroscopy
7.2.1 Basis of Quantitative Absorption Spectroscopy
The objective of quantitative absorption spectroscopy is to determine the concentration of analyte in a given sample solution. The determination is based on the measurement of the amount of light absorbed from a reference beam as it passes through the sample solution. In some cases the analyte may naturally absorb radiation in the UV-Vis range, such that the chemical nature of the analyte is not modified during the analysis. In other cases analytes that do not absorb radiation in the UV-Vis range are chemically modified during the analysis, converting them to a species that absorbs radiation of the appropriate wavelength. In either case the presence of analyte in the solution will affect the amount of radiation transmitted through the solution, and, hence, the relative transmittance or absorbance of the solution may be used as an index of analyte concentration.
-
T = transmittance
-
P 0 = radiant power of beam incident on absorption cell
-
P = radiant power of beam exiting the absorption cell
-
%T = percent transmittance
Absorbance is a convenient expression in that, under appropriate conditions, it is directly proportional to the concentration of the absorbing species in the solution. Note that based on these definitions for A and T, the absorbance of a solution is not simply unity minus the transmittance. In quantitative spectroscopy, the fraction of the incident beam that is not transmitted does not equal the solution’s absorbance (A).
-
A = absorbance
-
c = concentration of absorbing species
-
b = pathlength through solution (cm)
-
a = absorptivity
-
A and b = as in Eq. 7.4
-
ε = molar absorption coefficient
-
c = concentration in units of molarity
7.2.2 Deviations from Beer’s Law
It should never be assumed that Beer’s law is strictly obeyed. Indeed, there are several reasons for which the predicted linear relationship between absorbance and concentration may not be observed. In general, Beer’s law is applicable only to dilute solutions, up to approximately 10 mM for most analytes. The actual concentration at which the law becomes limiting will depend on the chemistry of the analyte. As analyte concentrations increase, the intermolecular distances in a given sample solution will decrease, eventually reaching a point at which neighboring molecules mutually affect the charge distribution of the other. This perturbation may significantly affect the ability of the analyte to capture photons of a given wavelength; that is, it may alter the analyte’s absorptivity (a). This causes the linear relationship between concentration and absorption to break down since the absorptivity term is the constant of proportionality in Beer’s law (assuming a constant pathlength, b). Other chemical processes also may result in deviations from Beer’s law, such as the reversible association-dissociation of analyte molecules or the ionization of a weak acid in an unbuffered solvent. In each of these cases, the predominant form of the analyte may change as the concentration is varied. If the different forms of the analyte (e.g., ionized versus neutral) have different absorptivities (a), then a linear relationship between concentration and absorbance will not be observed.
7.2.3 Procedural Considerations
The goal of many quantitative measurements is to determine the concentration of an analyte with optimum precision and accuracy, in a minimal amount of time, and at minimal cost. To accomplish this, it is essential that the analyst consider potential errors associated with each step in a particular assay. Potential sources of error for spectroscopic assays include inappropriate sample preparation techniques, inappropriate controls, instrumental noise, and errors associated with inappropriate conditions for absorbance measurements (such as extreme absorbance/transmittance readings).
Sample preparation schemes for absorbance measurements vary considerably. In the simplest case, the analyte-containing solution may be measured directly following homogenization and clarification. Except for special cases, homogenization is required prior to any analysis to ensure a representative sample. Clarification of samples is essential prior to taking absorbance readings in order to avoid the apparent absorption due to scattering of light by turbid solutions. The reference solution for samples in this simplest case will be the sample solvent, the solvent being water or an aqueous buffer in many cases. In more complex situations, the analyte to be quantified may need to be chemically modified prior to making absorbance measurements. In these cases, the analyte that does not absorb radiation in an appropriate spectral range is specifically modified, resulting in a species with absorption characteristics compatible with a given spectrophotometric measurement. Specific reactions such as these are used in many colorimetric assays that are based on the absorption of radiation in the Vis range. The reference solution for these assays is prepared by treating the sample solvent in a manner identical with that of the sample. The reference solution therefore will help to correct for any absorbance due to the modifying reagents themselves and not the modified analyte.
A sample-holding cell or cuvette should be chosen after the general spectral region to be used in a spectrophotometric measurement has been determined. Sample-holding cells vary in composition and dimensions. The sample-holding cell should be composed of a material that does not absorb radiation in the spectral region being used. Cells meeting this requirement for measurements in the UV range may be composed of quartz or fused silica. For the Vis range cells made of silicate glass are appropriate, and inexpensive plastic cells also are available for some applications. The dimensions of the cell will be important with respect to the amount of solution required for a measurement and with regard to the pathlength term used in Beer’s law. A typical absorption cell is 1 cm2 and approximately 4.5 cm long. The pathlength for this traditional cell is 1 cm, and the minimum volume of solution needed for standard absorption measurements is approximately 1.5 mL. Absorption cells with pathlengths ranging from 1 to 100 mm are commercially available. Narrow cells, approximately 4 mm in width, with optical pathlengths of 1 cm, are also available. These narrow cells are convenient for absorbance measurements when limiting amounts of solution are available, e.g., less than 1 mL.
The actual absorbance measurement is made by first calibrating the instrument for 0 % and then 100 % transmittance. The 0 % transmittance adjustment is made while the photodetector is screened from the incident radiation by means of an occluding shutter, mimicking infinite absorption. This adjustment sets the base level current or “dark current” to the appropriate level, such that the readout indicates zero. The 100 % transmittance adjustment then is made with the occluding shutter open and an appropriate reference cell/solution in the light path. The reference cell itself should be equivalent to the cell that contains the sample (i.e., a “matched” set of cells is used). In many cases, the same cell is used for both the sample and reference solutions. The reference cell generally is filled with solvent, that often being distilled/deionized water for aqueous systems. The 100 % T adjustment effectively sets T = 1 for the reference cell, which is equivalent to defining P 0 in Eq. 7.1 as equivalent to the radiant power of the beam exiting the reference cell. The 0 % T and 100 % T settings should be confirmed as necessary throughout the assay. The sample cell that contains analyte then is measured without changing the adjustments. The adjustments made with the reference cell will effectively set the instrument to give a sample readout in terms of Eq. 7.6. The readout for the sample solution will be between 0 and 100 % T. Most modern spectrophotometers allow the analyst to make readout measurements in either absorbance units or as percent transmittance. It is generally most convenient to make readings in absorbance units since, under optimum conditions, absorbance is directly proportional to concentration. When making measurements with an instrument that employs an analog swinging needle type of readout, it may be preferable to use the linear percent transmittance scale and then calculate the corresponding absorbance using Eq. 7.3. This is particularly true for measurements in which the percent transmittance is less than 20.
7.2.4 Calibration Curves
-
V s = volume of standard
-
V u = volume of unknown
-
V t = total volume
-
C s = concentration of standard
-
C u = concentration of unknown
-
k = proportionality constant (pathlength x absorptivity)
-
V s, V u, V t, C s, C u, and k = as in Eq. 7.8
7.2.5 Effect of Indiscriminant Instrumental Error on the Precision of Absorption Measurements
All spectrophotometric assays will have some level of indiscriminant error associated with the absorbance/transmittance measurement itself. Indiscriminant error of this type often is referred to as instrument noise. It is important that the assay be designed such that this source of error is minimized, the objective being to keep this source of error low relative to the variability associated with other aspects of the assay, such as sample preparation, subsampling, reagent handling, and so on. Indiscriminant instrumental error is observed with repeated measurements of a single homogeneous sample. The relative concentration uncertainty resulting from this error is not constant over the entire percent transmittance range (0–100 %). Measurements at intermediate transmittance values tend to have lower relative errors, thus greater relative precision, than measurements made at either very high or very low transmittance. Relative concentration uncertainty or relative error may be defined as S c/C, where S c is sample standard deviation and C is measured concentration. Relative concentration uncertainties of from 0.5 % to 1.5 % are to be expected for absorbance/transmittance measurements taken in the optimal range. The optimal range for absorbance measurements on simple, less expensive spectrophotometers is from approximately 0.2–0.8 absorbance units, or 15–65 % transmittance. On more sophisticated instruments, the range for optimum absorbance readings may be extended up to 1.5 or greater. To be safe, it is prudent to always make absorbance readings under conditions at which the absorbance of the analyte solution is less than 1.0. If there is an anticipated need to make measurements at absorbance readings greater than 1.0, then the relative precision of the spectrophotometer should be established experimentally by repetitive measurements of appropriate samples. Absorbance readings outside the optimal range of the instrument may be used, but the analyst must be prepared to account for the higher relative error associated with these extreme readings. When absorbance readings approach the limits of the instrumentation, then relatively large differences in analyte concentrations may not be detected.
7.2.6 Instrumentation
7.2.6.1 Light Source
Light sources used in spectrophotometers must continuously emit a strong band of radiation encompassing the entire wavelength range for which the instrument is designed. The power of the emitted radiation must be sufficient for adequate detector response, and it should not vary sharply with changes in wavelength or drift significantly over the experimental time scale. The most common radiation source for Vis spectrophotometers is the tungsten filament lamp. These lamps emit adequate radiation covering the wavelength region from 350 to 2,500 nm. Consequently, tungsten filament lamps also are employed in near-infrared spectroscopy. The most common radiation sources for measurements in the UV range are deuterium electrical-discharge lamps. These sources provide a continuous radiation spectrum from approximately 160 nm through 375 nm. These lamps employ quartz windows and should be used in conjunction with quartz sample holders, since glass significantly absorbs radiation below 350 nm.
7.2.6.2 Monochromator
The size of the wavelength range passing out of the exit slit of the monochromator is termed the bandwidth of the emitted radiation. Many spectrophotometers allow the analyst to adjust the size of the monochromator exit slit (and entrance slit) and, consequently, the bandwidth of the emitted radiation. Decreasing the exit slit width will decrease the associated bandwidth and the radiant power of the emitted beam. Conversely, further opening of the exit slit will result in a beam of greater radiant power but one that has a larger bandwidth. In some cases where resolution is critical, such as some qualitative work, the narrower slit width may be advised. However, in most quantitative work, a relatively open slit may be used since adsorption peaks in the UV-Vis range generally are broad relative to spectral bandwidths. Also, the signal-to-noise ratio associated with transmittance measurements is improved due to the higher radiant power of the measured beam.
Referring to Fig. 7.8, lines 1 and 2 represent rays of parallel monochromatic radiation that are in phase and that strike the grating surface at an angle i to the normal. Maximum constructive interference of this radiation is depicted as occurring at an angle r to the normal. At all other angles, the two rays will partially or completely cancel each other. Radiation of a different wavelength would show maximum constructive interference at a different angle to the normal. The wavelength dependence of the diffraction angle can be rationalized by considering the relative distance the photons of rays 1 and 2 travel and assuming that maximum constructive interference occurs when the waves associated with the photons are completely in phase. Referring to Fig. 7.8, prior to reflection, photon 2 travels a distance CD greater than photon 1. After reflection, photon 1 travels a distance AB greater than photon 2. Hence, the waves associated with photons 1 and 2 will remain in phase after reflection only if the net difference in the distance traveled is an integral multiple of their wavelength. Note that for a different angle r the distance AB would change and, consequently, the net distance CD–AB would be an integral multiple of a different wavelength. The net result is that the component wavelengths are each diffracted at their own unique angles r.
7.2.6.3 Detector
Photodiode detectors are now common in UV-Vis spectrophotometers. These are solid-state devices in which the light-induced electrical signal is a result of photons exciting electrons in the semiconductor materials from which they are fabricated, most commonly silicon. Spectrophotometers using photodiode detectors may contain a single diode detector or a linear array of diodes (diode array spectrophotometers). If a single photodiode detector is used, then the arrangement of components is generally as depicted in Fig. 7.6. If an array of photodiode detectors is used, then the light originating from the source typically passes into the sample prior to it being dispersed. The light transmitted through the sample is subsequently dispersed onto the diode array, with each diode measuring a narrow band of the resulting spectrum. This design allows one to simultaneously measure multiple wavelengths, allowing nearly instantaneous collection of an entire absorption spectrum. Diode-based detectors are generally reported to be more sensitive than phototubes but less sensitive than photomultiplier tubes.
7.2.6.4 Readout Device
The signal from the detector generally is amplified and then displayed in a usable form to the analyst. The final form in which the signal is displayed will depend on the complexity of the system. In the simplest case, the analog signal from the detector is displayed on an analog meter through the position of a needle on a meter face calibrated in percent transmission or absorbance. Analog readouts are adequate for most routine analytical purposes; however, analog meters are somewhat more difficult to read, and, hence, the resulting data are expected to have somewhat lower precision than that obtained on a digital readout (assuming the digital readout is given to enough places). Digital readouts express the signal as numbers on the face of a meter. In these cases, there is an obvious requirement for signal processing between the analog output of the detector and the final digital display. In virtually all cases, the signal processor is capable of presenting the final readout in terms of either absorbance or transmittance. Many of the newer instruments include microprocessors capable of more extensive data manipulations on the digitized signal. For example, the readouts of some spectrophotometers may be in concentration units, provided the instrument has been correctly calibrated with appropriate reference standards.
7.2.7 Instrument Design
7.2.8 Characteristics of UV-Vis Absorbing Species
Representative absorption maxima above 200 nm for select functional groups
Chromophore |
Example |
λ max a |
ε max b |
Resource material |
---|---|---|---|---|
Nonconjugated systems |
||||
R-CHO |
Acetaldehyde |
290 |
17 |
4 |
R2-CO |
Acetone |
279 |
15 |
4 |
R-COOH |
Acetic acid |
208 |
32 |
4 |
R-CONH2 |
Acetamide |
220 |
63 |
24 |
R-SH |
Mercaptoethane |
210 |
1,200 |
19 |
Conjugated systems |
||||
R2C=CR2 |
Ethylene |
<200 |
– |
24 |
R-CH=CH-CH=CH-R |
1,3 Butadiene |
217 |
21,000 |
24 |
R-CH=CH-CH=CH-CH=CH-R |
1,3,5 Hexatriene |
258 |
35,000 |
24 |
11 conjugated double bonds |
β-Carotene |
465 |
125,000 |
13 |
R2C=CH-CH=O |
Acrolein (2-propenal) |
210 |
11,500 |
24 |
315 |
14 |
24 |
||
HOOC-COOH |
Oxalic acid |
250 |
63 |
24 |
Aromatic compoundsc |
||||
C6H6 |
Benzene |
256 |
200 |
24 |
C6H5OH |
Phenol |
270 |
1,450 |
24 |
C8H7N |
Indole |
278 |
2,500 |
NIST databased |
The type of information contained in Table 7.2 will likely be useful in determining the feasibility of UV-Vis spectroscopy for specific applications. For example, it is helpful to know the absorption characteristics of carboxyl groups if one is considering the feasibility of using UV-Vis absorption spectroscopy as a detection method to monitor non-derivatized organic acids eluting from liquid chromatography columns. With respect to this particular organic acids question, the table indicates that organic acids are likely to absorb radiation in the range accessible to most UV-Vis detectors (>200 nm). However, the table also indicates that sensitivity of such a detection method is likely to be limited due to the low molar absorption coefficient of carboxyl groups at such wavelengths. This explains why high-performance liquid chromatography methods for organic acid quantification sometimes make use of UV-Vis-detectors tuned to ~210 nm (e.g., Resource Material 3) and why there are research efforts aimed at developing derivatization methods to enhance the sensitivity of UV-Vis-based methods for the quantification of organic acids (Resource Material 11).
The data of Table 7.2 also illustrate the effect of conjugation on electronic transitions. Increased conjugation leads to absorption maxima at longer wavelengths due to the associated decrease in the electronic energy spacing within a conjugated system (i.e., lower energy difference between the ground and excited state). The aromatic compounds included in the table were chosen due to their relevance to protein quantification: benzene/phenylalanine, phenol/tyrosine, and indole/tryptophan (Chap. 18, Sect. 18.5.1). The table indicates that typical proteins will have an absorption maximum at approximately 278 nm (high molar absorption coefficient of the indole side chain of tryptophan), as well as another peak at around 220 nm. This latter peak corresponds to the amide/peptide bonds along the backbone of the protein, the rationale being deduced from the data for the simple amide included in the table (i.e., acetamide).
7.3 Fluorescence Spectroscopy
Fluorescence spectroscopy is generally one to three orders of magnitude more sensitive than corresponding absorption spectroscopy. In fluorescence spectroscopy, the signal being measured is the electromagnetic radiation that is emitted from the analyte as it relaxes from an excited electronic energy level to its corresponding ground state. The analyte is originally activated to the higher energy level by the absorption of radiation in the UV or Vis range. The processes of activation and deactivation occur simultaneously during a fluorescence measurement. For each unique molecular system, there will be an optimum radiation wavelength for sample excitation and another, of longer wavelength, for monitoring fluorescence emission. The respective wavelengths for excitation and emission will depend on the chemistry of the system under study.
-
P F = radiant power of beam emitted from fluorescent cell
-
φ = constant of proportionality
-
P 0 and P = as in Eq. 7.1
Equation 7.17 is particularly useful because it emphasizes two important points that are valid for the conditions assumed when deriving the equation, particularly the assumption that analyte concentrations are kept relatively low. First, the fluorescent signal will be directly proportional to the analyte concentration, assuming other parameters are kept constant. This is very useful because a linear relationship between signal and analyte concentration simplifies data processing and assay troubleshooting. Second, the sensitivity of a fluorescent assay is proportional to P o, the power of the incident beam, the implication being that the sensitivity of a fluorescent assay may be modified by adjusting the source output.
7.4 Summary
UV and Vis absorption and fluorescence spectroscopy are used widely in food analysis. (See Chap. 8, Table 8.2, for a comparison of these types of spectroscopy, including their applications. The table also allows for comparison with other types of spectroscopy.) These techniques may be used for either qualitative or quantitative measurements. Qualitative measurements are based on the premise that each analyte has a unique set of energy spacings that will dictate its absorption/emission spectrum. Hence, qualitative assays generally are based on the analysis of the absorption or emission spectrum of the analyte. In contrast, quantitative assays most often are based on measuring the absorbance or fluorescence of the analyte solution at one wavelength. Quantitative absorption assays are based on the premise that the absorbance of the test solution will be a function of the solution’s analyte concentration.
Under optimum conditions, there is a direct linear relationship between a solution’s absorbance and its analyte concentration. The equation describing this linear relationship is known as Beer’s law. The applicability of Beer’s law to any given assay always should be verified experimentally by means of a calibration curve. The calibration curve should be established at the same time and under the same conditions that are used to measure the test solution. The analyte concentration of the test solution then should be estimated from the established calibration curve.
Molecular fluorescence methods are based on the measurement of radiation emitted from excited analyte molecules as they relax to lower energy levels. The analytes are raised to the excited state as a result of photon absorption. The processes of photon absorption and fluorescence emission occur simultaneously during the assay. Quantitative fluorescence assays are generally one to three orders of magnitude more sensitive than corresponding absorption assays. Like absorption assays, under optimal conditions there will be a direct linear relationship between the fluorescence intensity and the concentration of the analyte in the unknown solution. Most molecules do not fluoresce and, hence, cannot be assayed by fluorescence methods.
Instruments used for absorption and fluorescence methods have similar components, including a radiation source, wavelength selector(s), sample-holding cell(s), radiation detector(s), and a readout device.
7.5 Study Questions
- 1.
Why is it common to use absorbance values rather than transmittance values when doing quantitative UV-Vis spectroscopy?
- 2.
For a particular assay, the plot of absorbance vs. concentration is not linear; explain the possible reasons for this.
- 3.
What criteria should be used to choose an appropriate wavelength at which to make absorbance measurements, and why is that choice so important?
- 4.
In a particular assay, the absorbance reading on the spectrophotometer for one sample is 2.033 and for another sample 0.032. Would you trust these values? Why or why not?
- 5.
Explain the difference between electromagnetic radiation in the UV and Vis ranges. How does quantitative spectroscopy using the UV range differ from that using the Vis range?
- 6.
What is actually happening inside the spectrophotometer when the analyst “sets” the wavelength for a particular assay?
- 7.
Considering a typical spectrophotometer, what is the effect of decreasing the exit slit width of the monochromator on the light incident to the sample?
- 8.
Describe the similarities and differences between a phototube and a photomultiplier tube. What is the advantage of one over the other?
- 9.
Your lab has been using an old single-beam spectrophotometer that must now be replaced by a new spectrophotometer. You obtain sales literature that describes single-beam and double-beam instruments. What are the basic differences between a single-beam and a double-beam spectrophotometer, and what are the advantages and disadvantages of each?
- 10.
Explain the similarities and differences between UV-Vis spectroscopy and fluorescence spectroscopy with regard to instrumentation and principles involved. What is the advantage of using fluorescence spectroscopy?
7.6 Practice Problems
- 1.
A particular food coloring has a molar absorption coefficient of 3.8 × 103 cm−1 M −1 at 510 nm.
- (a)
What will be the absorbance of a 2 × 10−4 M solution in a 1-cm cuvette at 510 nm?
- (b)
What will be the percent transmittance of the solution in (a)?
- (a)
- 2.
- (a)
You measure the percent transmittance of a solution containing chromophore X at 400 nm in a 1-cm pathlength cuvette and find it to be 50 %. What is the absorbance of this solution?
- (b)
What is the molar absorption coefficient of chromophore X if the concentration of X in the solution measured in question 2a is 0.5 mM?
- (c)
What is the concentration range of chromophore X that can be assayed if, when using a sample cell of pathlength 1, you are required to keep the absorbance between 0.2 and 0.8?
- (a)
- 3.
What is the concentration of compound Y in an unknown solution if the solution has an absorbance of 0.846 in a glass cuvette with a pathlength of 0.2 cm? The absorptivity of compound Y is 54.2 cm−1 (mg/mL)−1 under the conditions used for the absorption measurement.
- 4.
- (a)
What is the molar absorption coefficient of compound Z at 295 and 348 nm, given the absorption spectrum shown in Fig. 7.14 (which was obtained using a UV-Vis spectrophotometer and a 1 mM solution of compound Z in a sample cell with a pathlength of 1 cm)?
- (b)
Assume you decide to make quantitative measurements of the amount of compound Z in different solutions. Based on the above spectrum, which wavelength will you use for your measurements? Give two reasons why this is the optimum wavelength.
- (a)
Answers
- 1.
(a) = 0.76, (b) = 17.4
This problem requires a knowledge of the relationship between absorbance and transmittance and the ability to work with Beer’s law.
Given: molar absorption coefficient = 3.8 × 103 cm−1 M −1 - 2.
(a) = 301, (b) = 602 cm−1 M −1, (c) = 0.33 × 10−3 M to 1.33 × 10−3 M
This problem again requires knowledge of the relationship between absorbance and transmittance and the manipulation of Beer’s law. Care must be taken in working with the appropriate concentration units.- (a)
T = 0.5
- (b)
Given that the solution in part (a) is 0.5 mM (equivalent to 5 × 10−4 M)
Rearranging Beer’s law: ε = A/(bc) - (c)
To answer the problem, find the concentration that will give an absorbance of 0.200 (lower limit) and the concentration that will give an absorbance of 0.800 (upper limit). In both cases, use Beer’s law to determine the appropriate concentrations:
where:-
Lowest concentration = 0.2/[(602 cm−1 M −1)(1 cm)] = 3.3 × 10−4 M (i.e., 0.33 mM)
-
Highest concentration = 0.8/[(602 cm−1 M −1)(1 cm)] = 1.3 × 10−3 M (i.e., 1.33 mM)
-
- (a)
- 3.
0.078 mg/mL
This problem illustrates (1) that concentration need not be expressed in units of molarity and (2) that the pathlength of the cuvette must be considered when applying Beer’s law. In the present problem the analyte concentration is given in mg/mL: thus, the absorptivity must be in analogous units: - 4.
(a) = 860 at 295 nm, 60 at 348 nm; (b) = 295 nm; optimum sensitivity and more likely to adhere to Beer’s law.
This problem presents the common situation in which one wants to use absorbance spectroscopy for quantitative measurements but is unsure what wavelength to choose for the measurements. Furthermore, the absorptivity of the analyte at the different wavelengths of interest is unknown. A relatively simple way to obtain the necessary information is to determine the absorption spectrum of the analyte at a known concentration.- (a)
The arrows on the provided spectrum indicate the points on the spectrum corresponding to 295 and 348 nm. The problem notes that the absorption spectrum was obtained using a 1 mM solution (i.e., 1 × 10−3 M solution) of the analyte and that the pathlength of the cuvette was 1 cm. The answer to the problem is thus determined by taking the absorbance of the analyte at the two wavelengths in question and then plugging the appropriate data into Beer’s law. It is somewhat difficult to get an exact absorbance reading from the presented spectrum, but we can estimate that the absorbance of the 1 mM solution is ~0.86 at 295 nm and ~0.06 at 348 nm.
-
At 295 nm ε = 0.86/[(1 cm) (.001 M)] = 860 cm−1 M −1
-
At 348 nm ε = 0.06/[(1 cm) (.001 M)] = 60 cm−1 M −1
-
- (b)
In general, analysts strive to obtain maximum sensitivity for their assays, where sensitivity refers to the change in assay signal per unit change in analyte concentration (the assay signal in this case is absorbance). The absorbance values for the analyte at the different wavelengths, taken from the absorption spectrum, and/or the relative absorptivity values for the analyte at the different wavelengths, provide a good approximation of the relative sensitivity of the assay at different wavelengths (it is an approximation because we have not determined the variability/precision of the measurements at the different wavelengths). It can be seen from the given spectrum that absorbance “peaks” were at ~298 and ~370 nm. The sensitivity of the assay, relative to neighboring wavelengths, is expected to be maximum at these absorbance peaks. The peak at 295 nm is significantly higher than that at 370 nm, so the sensitivity of the assay is expected to be significantly higher at 295 nm. Thus, this would be the optimum wavelength to use for the assay. A second reason to choose 295 nm is because it appears to be in the middle of the “peak,” and, thus, small changes in wavelength due to instrumental/operator limitations are not expected to appreciably change the absorptivity values. Therefore, the assay is more likely to adhere to Beer’s law.
There are situations in which an analyst may choose to not use the wavelength corresponding to an overall maximum absorbance. For example, if there are known to be interfering compounds that absorb at 295 nm, then an analyst may choose to do take absorbance measurements at 370 nm.
- (a)
Open Access This chapter is licensed under the terms of the Creative Commons Attribution-NonCommercial 2.5 International License (http://creativecommons.org/licenses/by-nc/2.5/), which permits any noncommercial use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license and indicate if changes were made.
The images or other third party material in this chapter are included in the chapter's Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the chapter's Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.